ASP Global, LLC. dba Anatomy Supply Partners, LLC.
FDA classification date
Oct 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Source facts
Action
Starting on 5//5/2025 customers were informed that there had been reports of falsely elevated blood lead levels associated with the use of affected micro capillary tubes for lead testing when used with Magellan LeadCare Testing Systems.
On 9/8/2025, recall notices were emailed to customers who were asked to do the following:
Discontinue use of the recalled products immediately.
b) Quarantine any remaining inventory to prevent further use.
c) Dispose remaining inventory in accordance with local regulations or notify firm at
FieldActions@aspglobal.com for return authorization and disposal assistance.
d) Complete and return the enclosed Recall Response Form to confirm receipt and actions
to FieldActions@aspglobal.com
Distributors are asked to notify their customers of the recall or to forward their customer information to the recalling firm so they might notify these customers.
Initiated
2025-05-05
Root cause
Component design/selection
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Recall number
Z-0372-2026
Recall status
Open, Classified
CDRH record ID
215914
FDA classified
2025-10-27
Recall event number
97551