Open original FDA record ↗Recall record
- FDA identifier
- Z-0374-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Nov 26, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.
Source facts
- Action
- Siemens Healthcare sent a Field Safety Notice dated September 3, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the Field Safety Notice to any new users and provide the new owner information to Siemens Healthcare. Customers were asked to send back the attached form confirming receipt of the Field Safety Notice. Customers with questions should call 610-850-4634. For questions regarding this recall call 610-219-6300.
- Initiated
- 2013-09-03
- Root cause
- Software design
- Terminated
- 2015-02-24
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada.
- Recall number
- Z-0374-2014
- Recall status
- Terminated
- CDRH record ID
- 121793
- FDA classified
- 2013-11-26
- Recall event number
- 66231