Open original FDA record ↗Recall record
- FDA identifier
- Z-0374-2016
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Dec 4, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sp02 and/or Non Invasive Blood Pressure (NBP) alarms may become disabled without visual notification
Source facts
- Action
- The firm, Philips Healthcare, issued a "URGENT-Field Safety Notice" on 12/19/14 to inform customers of the issue, identify details of the units affected, instructs on actions to be taken by the customer and identify what action Philips plans to take to remedy the issue. Philips will provide a software correction free of charge. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice-To prevent this issue from occurring, customers/users should: 1) Disable the Change Bed Label field in Patient Management. This is accomplished by changing the feature from Optional to Hidden 2) All patient transfers must be performed using the Transfer key in the Manage Patient application. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment. If you need any further information or support concerning this issue, please contact your local Philips Healthcare Customer Service representative or the Customer Care Solutions Center at (800) 722-9377 with questions or concerns about this Device Correction.
- Initiated
- 2014-12-19
- Root cause
- Software Design Change
- Terminated
- 2020-09-21
- Distribution
- Worldwide Distribution: US (nationwide) and countries: AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, HUNGARY, ICELAND, INDIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, LUXEMBOURG, MOROCCO, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Recall number
- Z-0374-2016
- Recall status
- Terminated
- CDRH record ID
- 132093
- FDA classified
- 2015-12-04
- Recall event number
- 70000