Z-0375-2017: Philips Healthcare System X-ray Tomography Computed :

Recall record Z-0375-2017 for Philips Electronics North America Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0375-2017
Company in FDA record
Philips Electronics North America Corporation
FDA record creation date
Nov 22, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered the patient support tabletop may fail to move in the horizontal direction when it is commanded to do so during CT scans due to a potential mechanical linkage failure.

Source facts

Action
Philips planned action: 1. Philips issued a customer notification letter on October 13, 2016. The letter includes a statement that the corrections will be made free of charge. 2. Philips will initiate a field corrective action to address the issue. 3. Philips will contact the customer and schedule an inspection to determine if the unit requires repair. 4. If repair is required, the couch horizontal linkage assembly will be replaced. For further questions please call (978) 659-3000.
Initiated
2016-03-15
Root cause
Radiation Control for Health and Safety Act
Terminated
2020-09-09
Distribution
USA(nationwide) Distribution to the states of : AZ, NY, GA, UT, TN, TX, AR, NV, Il, FL, MS, LA, MD, MO, MA, OK, KY, WY, SC, ND, NE, DE, MI, WA, CT, RI, KS, OR, IN, PA, CA, NJ, NC, MN, AL and IA.
Recall number
Z-0375-2017
Recall status
Terminated
CDRH record ID
150772
FDA record created
2016-11-22
Recall event number
75551

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗