Labeling Error: Some cartridges may be mislabeled as "Xpert EV" instead of "Xpert GBS". (The information embedded in the barcode label on the cartridge is correct for Xpert GBS and brings up the GBS assay definition file)
Source facts
Action
Consignees were sent a Correction and Removal letter on 09/04/2007, notifying them that there was a labeling error with some of the cartridges labels indicating Xpert EV instead of Xpert GBS on the cartridge label. They are asked to discontinue using it immediately and return to Cepheid.
Initiated
2007-09-05
Root cause
Release of Material/Component prior to receiving test results
Terminated
2008-04-04
Distribution
Worldwide Distribution
Recall number
Z-0376-2008
Recall status
Terminated
CDRH record ID
65632
FDA classified
2008-01-12
Recall event number
45494
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.