Open original FDA record ↗Recall record
- FDA identifier
- Z-0376-2016
- Company in FDA record
- Windstone Medical Packaging, Inc.
- FDA record creation date
- Dec 4, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AMS 4674 Custom Pack is recalled because this kit containing U-SDN 2 Part Ultra Modified Seldinger Needle which is recalled by supplier.
Source facts
- Action
- Aligned Medical Solutions (AMS) sent the Procedure Products -Urgent Medical Device Recall Letter, dated November 02, 2015 to consignees. This letter listed the items affected by recall as: AMS 4674 Custom Pack and AMS 4675 Angio Pack. Both direct account and sub account are informed that the part # NS/37 -0387 01 U-SDN 2 Part Ultra Modified Seldinger Needle (introducer needle) is recalled because some was unable to accept guide wires up to .038" as intended. The inner diameter was found to be out of specification in some cases. All other components in these two AMS kits are not affected by this recall. Customers with questions about this recall can call Vicki Davis at 321-527-7714 or send email to vdavis@alignedmedicalsolutions.com.
- Initiated
- 2015-11-02
- Root cause
- Process control
- Terminated
- 2016-08-17
- Distribution
- Distributed in AL and VA.
- Recall number
- Z-0376-2016
- Recall status
- Terminated
- CDRH record ID
- 141641
- FDA record created
- 2015-12-04
- Recall event number
- 72604