Z-0377-2025: Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4482; use in Peritoneal Di

Recall record Z-0377-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0377-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Nov 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

Source facts

Action
Baxter issued an "Important Medical Device Correction" notice to consignees beginning on 10/21/2024 via letter. In February 2025, Baxter Kidney Care segment launched a standalone company, Vantive. Vantive issued an updated notice to consignees beginning on 2/19/25 via letter. The notice explained the issue, hazard involved, and requested the following actions be taken: For patients older than 6 months of age, healthcare providers are advised to continue using both the peroxide-cured silicone tubing version and/or the platinum-cured silicone tubing version of the MiniCap Extended Life PD transfer sets, as neither is anticipated to present safety risks related to PCB/PCBAs. 2. For patients younger than 6 months: a. Healthcare providers should prioritize the use of platinum-cured silicone tubing sets or other alternatives that may be available. b. If no alternatives are available, healthcare providers should continue to connect the peroxide-cured silicone tubing sets and prioritize the use of shorter transfer sets. c. If the patient currently has the peroxide-cured silicone tubing, they do not need to have it replaced early, as data demonstrates that PCBA levels decrease over treatment time. 3. If you received this communication directly from Vantive, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. 4. If you purchased this product from a distributor, please note that responding via the Vantive customer portal is not applicable. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers.
Initiated
2024-10-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide Distribution (US Nationwide distribution)
Recall number
Z-0377-2025
Recall status
Open, Classified
CDRH record ID
210534
FDA classified
2024-11-20
Recall event number
95525

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗