Z-0378-06: GEM PREMIER 3000

Recall record Z-0378-06 for Instrumentation Laboratory Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0378-06
Company in FDA record
Instrumentation Laboratory Co.
FDA classification date
Jan 11, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software error may report higher results for glucose and lactate

Source facts

Action
Instrumentation Laboratory notified customers by letter dated December 2005 on 12/16/2005. Customers were notifed if the error message occurs the lab should diable the glucose and lactate channels or remove cartridge. A software upgrade will be released in early 2006.
Initiated
2005-12-16
Root cause
Other
Terminated
2007-05-21
Distribution
Nationwide Foreign: Canada
Recall number
Z-0378-06
Recall status
Terminated
CDRH record ID
43338
FDA classified
2006-01-11
Recall event number
34249

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗