Reason stated in the source
Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
Source facts
- Action
- Roche sent an Urgent Medical Device Correction (UMDC) sent via UPS Ground (receipt required) dated July 6, 2017, and an amended UMDC on August 31 ,2017. All consignees will be notified by Urgent Medical Device Correction sent via UPS Ground (receipt required). Customers/consignees were instructed the following: If the Tip/Cup pick up error alarm is issued then followed by the Abnormal L2-line Movement alarm on your MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer, after which the system goes into Sampling Stop, follow the steps in this UMDC and the appropriate How to Identify Potentially Mismatched Sample Results within the enclosed document. For further questions, please call (317) 521-4343.
- Initiated
- 2017-06-29
- Root cause
- Software design
- Terminated
- 2020-01-24
- Distribution
- USA (nationwide) Distribution
- Recall number
- Z-0378-2018
- Recall status
- Terminated
- CDRH record ID
- 160292
- FDA classified
- 2018-01-16
- Recall event number
- 78760