Skip to content

Public FDA record

Z-0379-2013: Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. In

Recall record Z-0379-2013 for Microgenics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0379-2013
Company in FDA record
Microgenics Corporation
FDA classification date
Nov 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.

Source facts

Action
Microgenics Corporation sent a Recall Notification letter dated August 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed that immediate attention is required and identify the two affected constituents. The letter contains a list of the affected product with their catalog number, lot number and expiration date and request that the CK-MB and Insulin use be discontinued. The remaining components of the product may be used. Evaluate quality controls according to your laboratory procedures. Retain a copy of the letter for your laboratory records. If the product was forwarded to another laboratory, provide them with a copy of the letter. Complete the attached Recall Response form and return within 5 days. Questions may be directed to 1-800-232-3342.
Initiated
2012-08-20
Root cause
Nonconforming Material/Component
Terminated
2013-03-04
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Germany, Canada and Australia.
Recall number
Z-0379-2013
Recall status
Terminated
CDRH record ID
112207
FDA classified
2012-11-19
Recall event number
62987

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗