Skip to content

Public FDA record

Z-0379-2018: Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components m

Recall record Z-0379-2018 for Carl Zeiss Metrology Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0379-2018
Company in FDA record
Carl Zeiss Metrology Inc
FDA record creation date
Jan 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Source facts

Action
Zeiss sent an Safety Notice Letter dated September 28, 2017, to the customers advising them of the recall. The affected device was identified as well as the issue involved. Customers are to complete and return the confirmation- waiver form. During the next scheduled service event, a ZEISS field service technician will update the Metrotom operating system. The corrections are to be made free of charge, and will be completed by March 2018. CDRH approves the CAP subject to the following conditions: Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. For further questions, please call (763) 203-4267.
Initiated
2017-11-14
Root cause
Radiation Control for Health and Safety Act
Distribution
Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
Recall number
Z-0379-2018
Recall status
Open, Classified
CDRH record ID
160805
FDA record created
2018-01-26
Recall event number
78885

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗