Open original FDA record ↗Recall record
- FDA identifier
- Z-0379-2018
- Company in FDA record
- Carl Zeiss Metrology Inc
- FDA record creation date
- Jan 26, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
Source facts
- Action
- Zeiss sent an Safety Notice Letter dated September 28, 2017, to the customers advising them of the recall. The affected device was identified as well as the issue involved. Customers are to complete and return the confirmation- waiver form. During the next scheduled service event, a ZEISS field service technician will update the Metrotom operating system. The corrections are to be made free of charge, and will be completed by March 2018. CDRH approves the CAP subject to the following conditions: Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. For further questions, please call (763) 203-4267.
- Initiated
- 2017-11-14
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
- Recall number
- Z-0379-2018
- Recall status
- Open, Classified
- CDRH record ID
- 160805
- FDA record created
- 2018-01-26
- Recall event number
- 78885