A software anomaly exists in the 5.20.1038 NicoletOne LTM, Sleep, nEEG, vEEG and ICU Monitor systems using the M, and/or C series amplifiers. If this anomaly occurs, the system will display the data for channel one in all channels on NicoletOne LTM, Sleep, nEEG, vEEG or ICU Monitor systems.
Source facts
Action
An Advisory Notice, dated August 25, 2006, was sent to all affected customers. The notice described the issues, states to discontinue use of 5.20.1038 software until updated 5.21.1039c software is installed.
Initiated
2006-08-25
Root cause
Other
Terminated
2011-12-10
Distribution
Worldwide, including USA, Belgium, Chile, Denmark, Italy, Japan, Poland, and Spain.
Recall number
Z-0380-2007
Recall status
Terminated
CDRH record ID
48241
FDA classified
2007-02-15
Recall event number
36241