Z-0380-2011: Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084500800, Manufactured by Sorin Group USA, Inc. 14

Recall record Z-0380-2011 for Sorin Group USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0380-2011
Company in FDA record
Sorin Group USA, Inc.
FDA classification date
Nov 15, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Defective product packaging may compromise product sterility.

Source facts

Action
Sorin Group sent an "URGENT: FIELD SAFETY NOTICE" letter dated October 7, 2010, to all customers. The letter described the product, problem and action to be taken by customer. The customers were instructed to check all affected inventory to exam for holes. Any damaged products were to be returned to Sorin. If no damage was found, product could be used. The customers were also instructed to contact Sorin Cardiopulmonary Customer Service at 1-800-650-2623 to return product and receive replacement; they were to share this information with all personnel who may use this product; and to complete and return the attached Customer Response Form via the self-addressed prepaid envelope provided or fax to (303) 467-6502 as soon as possible. For further information, contact 1-800-650-2623.
Initiated
2010-10-07
Root cause
Packaging process control
Terminated
2010-12-17
Distribution
Nationwide distribution: CA, FL, IL, IN, KS, MD, MO, NE, NJ, NM, TN, TX, and VA.
Recall number
Z-0380-2011
Recall status
Terminated
CDRH record ID
95267
FDA classified
2010-11-15
Recall event number
57052

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗