For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
Source facts
Action
Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.
Initiated
2017-07-17
Root cause
Device Design
Terminated
2018-08-29
Distribution
US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
Recall number
Z-0380-2018
Recall status
Terminated
CDRH record ID
160827
FDA record created
2018-01-17
Recall event number
78900