Skip to content

Public FDA record

Z-0380-2019: Liquid Cardiac Control, CQ5051

Recall record Z-0380-2019 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0380-2019
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Nov 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.

Source facts

Action
On October 2, 2018, the firm notified distributors of the recall via Expanded Urgent Medical Device Correction email. Customers were informed that the recall was related to previous recall REC334 on June 8, 2018. The firm has confirmed that Liquid Cardiac Controls CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials. Customers were advised to take the following actions: - Discontinue use of these products for the Quality Control (QC) monitoring of Troponin T assays. - Discuss the contents of this notice with your Medical Director. - Update kits with revised IFUs excluding Troponin T values and the attached important notice to prevent further use of the device in the QC of Troponin T assays. - Complete and return the vigilance response section of this form to technical.services@randox.com within five working days. - Contact your local Randox sales representative for alternative product details. You may contact the firm at technical.services@randox.com.
Initiated
2018-10-02
Root cause
Under Investigation by firm
Terminated
2019-05-30
Distribution
Distributed to IN, NY, PR, and WV.
Recall number
Z-0380-2019
Recall status
Terminated
CDRH record ID
167599
FDA record created
2018-11-03
Recall event number
81159

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗