Open original FDA record ↗Recall record
- FDA identifier
- Z-0381-2008
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Dec 1, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
Source facts
- Action
- The firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge.
- Initiated
- 2007-03-23
- Root cause
- Device Design
- Terminated
- 2011-02-11
- Distribution
- Worldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, Italy, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, China, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, Ukraine, UK.
- Recall number
- Z-0381-2008
- Recall status
- Terminated
- CDRH record ID
- 66447
- FDA classified
- 2007-12-01
- Recall event number
- 38345