Open original FDA record ↗Recall record
- FDA identifier
- Z-0381-2022
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Dec 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Manufacturing instructions did not include the directions to not retest the pneumatic kit component of the ROTAPRO console upon failure, an internal nonconformity.
Source facts
- Action
- The recalling firm issued letters dated 11/8/2021 on the same date and flagged "Urgent Medical Device Removal - Immediate Action Required." It explains the reason for recall, the risk to health, and provides removal instructions. The instructions say to immediately discontinue use of and segregate the affected product. A Reply Verification Tracking Form is enclosed for immediate completion and return. The customer will be contacted by BSC and provided a Returned Goods Authorization for return of the device.
- Initiated
- 2021-11-08
- Root cause
- Employee error
- Terminated
- 2023-05-02
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of FL, NV, NY, OH, PA, and WA. The countries of Germany.
- Recall number
- Z-0381-2022
- Recall status
- Terminated
- CDRH record ID
- 190351
- FDA record created
- 2021-12-15
- Recall event number
- 89034