Skip to content

Public FDA record

Z-0382-05: Medication Bottle (U22-3), a component of the MicroAir Vibrating Mesh Nebulizer, Model NE-U22V/NE-22VAC;

Recall record Z-0382-05 for Omron Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0382-05
Company in FDA record
Omron Healthcare, Inc.
FDA classification date
Jan 5, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A potential failure of the vibrating mechanism of the medication bottle which causes a failure to transmit vibration and results in no nebulization.

Source facts

Action
Omron sent 'Removal Notices' dated 11/2/04 to all of their direct accounts receiving the nebulizers during the affected time frame. The accounts were informed of the potential failure of the vibrating mechanism, and listed the serial numbers of the nebulizers that may contain the affected medication bottles. Medication bottles with either a blue dot near the bottle lever or a blue O-ring around the body of the bottle were manufactured after corrections were made and are not subject to the recall. Affected product should be returned to Omron for replacement. Any questions were directed to Customer Service at 800-231-4030.
Initiated
2004-11-02
Root cause
Other
Terminated
2007-06-26
Distribution
Nationwide
Recall number
Z-0382-05
Recall status
Terminated
CDRH record ID
35735
FDA classified
2005-01-05
Recall event number
30363

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗