Open original FDA record ↗Recall record
- FDA identifier
- Z-0382-05
- Company in FDA record
- Omron Healthcare, Inc.
- FDA classification date
- Jan 5, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A potential failure of the vibrating mechanism of the medication bottle which causes a failure to transmit vibration and results in no nebulization.
Source facts
- Action
- Omron sent 'Removal Notices' dated 11/2/04 to all of their direct accounts receiving the nebulizers during the affected time frame. The accounts were informed of the potential failure of the vibrating mechanism, and listed the serial numbers of the nebulizers that may contain the affected medication bottles. Medication bottles with either a blue dot near the bottle lever or a blue O-ring around the body of the bottle were manufactured after corrections were made and are not subject to the recall. Affected product should be returned to Omron for replacement. Any questions were directed to Customer Service at 800-231-4030.
- Initiated
- 2004-11-02
- Root cause
- Other
- Terminated
- 2007-06-26
- Distribution
- Nationwide
- Recall number
- Z-0382-05
- Recall status
- Terminated
- CDRH record ID
- 35735
- FDA classified
- 2005-01-05
- Recall event number
- 30363