Skip to content

Public FDA record

Z-0382-2009: ConMed Linvatec Hall Surgical Blades, Sterile ST OSC BLD, Catalog # 00507114500, 00507113200, 0050711740

Recall record Z-0382-2009 for Linvatec Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0382-2009
Company in FDA record
Linvatec Corp.
FDA classification date
Jan 15, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Sterility compromised. There is a possibility that certain catalog number and lots of ConMed Linvatec Hall Surgical Blades, may have a hole(s) in the corner(s) of the blister package(tray), thereby potentially compromising the sterility of the devices.

Source facts

Action
An Urgent Medical Device Recall letter was issued. Customers must check their facility's inventory immediately for the catalog and lot numbers in the attached list. They also need to complete and sign the enclosed Reply Form and mail or fax the Reply Form to ConMed Linvatec using the enclosed self addressed postage paid envelope or fax 727-319-5701. Customers have an option to either re-sterilize the products using the re-sterilization instructions in the letter or they can return the devices to ConMed Linvatec for a credit using the enclosed preprinted FedEx Air Bill with ConMed Linvatec's Bill Number and the return goods number. Recall is being conducted to the distributor and end user level.
Initiated
2008-02-21
Root cause
Process control
Terminated
2009-01-15
Distribution
Worldwide Distribution
Recall number
Z-0382-2009
Recall status
Terminated
CDRH record ID
70449
FDA classified
2009-01-15
Recall event number
47848

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.