Skip to content

Public FDA record

Z-0382-2012: CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is ind

Recall record Z-0382-2012 for Accuray Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0382-2012
Company in FDA record
Accuray Incorporated
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.

Source facts

Action
Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )
Initiated
2011-11-07
Root cause
Labeling design
Terminated
2012-02-27
Distribution
Worldwide Distribution -- USA ( nationwide ) including the states of FL, TN, IL, NJ, CO, PA, MD, CA, MS, MN, SC, WA, OK, GA, AZ, and TX., and the countries of France and United Kingdom.
Recall number
Z-0382-2012
Recall status
Terminated
CDRH record ID
105406
FDA classified
2012-01-11
Recall event number
60437

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗