Open original FDA record ↗Recall record
- FDA identifier
- Z-0382-2012
- Company in FDA record
- Accuray Incorporated
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.
Source facts
- Action
- Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )
- Initiated
- 2011-11-07
- Root cause
- Labeling design
- Terminated
- 2012-02-27
- Distribution
- Worldwide Distribution -- USA ( nationwide ) including the states of FL, TN, IL, NJ, CO, PA, MD, CA, MS, MN, SC, WA, OK, GA, AZ, and TX., and the countries of France and United Kingdom.
- Recall number
- Z-0382-2012
- Recall status
- Terminated
- CDRH record ID
- 105406
- FDA classified
- 2012-01-11
- Recall event number
- 60437