Skip to content

Public FDA record

Z-0382-2013: SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-

Recall record Z-0382-2013 for Stryker Spine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0382-2013
Company in FDA record
Stryker Spine
FDA classification date
Nov 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Source facts

Action
Stryker Spine sent an Urgent Medical Device Removal Notification letter/Product Accountability Forms dated August 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 201-760-8150.
Initiated
2012-08-09
Root cause
Nonconforming Material/Component
Terminated
2015-09-30
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
Recall number
Z-0382-2013
Recall status
Terminated
CDRH record ID
113780
FDA classified
2012-11-19
Recall event number
63386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗