Open original FDA record ↗Recall record
- FDA identifier
- Z-0383-2018
- Company in FDA record
- TEI Biosciences
- FDA record creation date
- Jan 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10 contained extra pages and/or missing pages: " 22 IFUs contained duplicate pages 15 - 26 & 51 - 62 and were also missing pages 27 - 50 " 3 IFUs were missing pages 27 - 50 " 1 IFU was missing pages 1 - 2 & 75 - 76 " 1 IFU was missing pages 15 - 62 While pages 27 50, 75 76, and 15 62 contain instructions written in a language other than English, pages 1 2 contain information regarding the indications for use, contraindications, device description, warning and precautions, and potential complications of the device.
Source facts
- Action
- Integra LifeSciences Corporation notified (all U.S.) customers via Letter/Medical Device Correction Field Safety Notice dated June 15, 2017. Letter delivered by courier service, facsimile or email July 15, 2017.
- Initiated
- 2017-06-15
- Root cause
- Error in labeling
- Terminated
- 2018-03-15
- Distribution
- U.S.
- Recall number
- Z-0383-2018
- Recall status
- Terminated
- CDRH record ID
- 160543
- FDA record created
- 2018-01-17
- Recall event number