Skip to content

Public FDA record

Z-0384-04: Products are various dental surgical drills packed in a sealed tray and labeled as sterile. The tray is

Recall record Z-0384-04 for Implant Innovations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0384-04
Company in FDA record
Implant Innovations, Inc.
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The tray lid seals may be open thereby compromising the sterility barriers for the recalled lots.

Source facts

Action
The firm notified international distributors on 12/2/2003 via e-mail and also a recall notification letter was faxed to them on 12/2/2003. They were instructed to subrecall to the health professional level. Also on 12/2/2003 all domestic accounts (health practioners) were telephoned and then faxed a recall notification letter. Each account was instructed to check their inventory, respond back by fax and to return recalled product to Implant Innovations Inc.
Initiated
2003-12-02
Root cause
Other
Terminated
2009-10-05
Distribution
Product was distributed to 45 Dentist/Surgeons nationwide in the US and also to 13 international Distributors in Canada, Sweden, Germany, Colombia, Spain, Switzerland, Australia, Japan, Austria, France, Poland, England and Greece.
Recall number
Z-0384-04
Recall status
Terminated
CDRH record ID
31189
FDA classified
2004-07-20
Recall event number
28092

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗