- FDA identifier
- Z-0385-05
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Jan 8, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
Source facts
- Action
- Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning the return process were directed to NCC at 1-866-300-5085. Any medical or technical questions were directed to Baxter at 1-800-422-9837.
- Initiated
- 2004-11-19
- Root cause
- Other
- Terminated
- 2006-06-28
- Distribution
- Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland
- Recall number
- Z-0385-05
- Recall status
- Terminated
- CDRH record ID
- 36017
- FDA classified
- 2005-01-08
- Recall event number
- 30506