Open original FDA record ↗Recall record
- FDA identifier
- Z-0385-2019
- Company in FDA record
- CareFusion 303, Inc.
- FDA record creation date
- Nov 4, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaints where users were unable to prime the administration set.
Source facts
- Action
- A "Medical Device Safety Notification" Letter dated April 20, 2018 was to all affected customers, stating that the firm has become aware of issues with infusion pump administration sets. Customers are experiencing difficultly or failing to prime the pump administrative sets. Infusion pump administration sets, various models with check valves manufactured during January 2017 through March 2018. Issue: The Infusion pump administration sets require priming prior to patient use as described in the directions for use. It is during the priming process that this issue can be observed. During priming, when attempting to fill the drip chamber, the drops into the drip chamber will not be observed once the roller clamp is opened. This issue is highly detectable. Steps for detecting this issue during priming: -Open administrative set package, remove set and close roller clamp. -Insert administrative set spike into the IV bag or bottle and hang container 20 inches above the Pump module. -Fill drip chamber to 2/3 full. -Open the roller clamp. -Drips into the drip chamber will be very slow or absent. Potential Risk: An occlusion in the administration set line can prevent the ability to prime the set, rendering the administration set unusable. The inability to prime can contribute to the delay of the start of the infusion. There have been no reports of patient harm as a result of this issue. Recommended Steps for Users: You can continue to use the product as this issue is highly detectable. If during priming you experience a difficultly or inability to prime, please notify CareFusion Customer Advocacy at 888-812-3266 or customerfeedback@BD.com. Meantime, BD has the following options if you experience this issue: In some instances clinicians may be able to increase the pressure to assist the check valve by gently squeezing the IV bag following these steps: 1. Begin by removing the IV set from the package and close the roller clamp. 2. Spike the bag of IV flu
- Initiated
- 2018-04-20
- Root cause
- Process control
- Terminated
- 2021-02-10
- Distribution
- Worldwide Distribution - US Nationwide including Guam International: Canada, Australia, New Zealand, EU, Middel East, South Africa, Okinawa, Saipan, Buenos Aires, Lima, Selangor, Hong Kong, Taiwan, and Budapest
- Recall number
- Z-0385-2019
- Recall status
- Terminated
- CDRH record ID
- 167114
- FDA record created
- 2018-11-04
- Recall event number
- 80956