Open original FDA record ↗Recall record
- FDA identifier
- Z-0386-2010
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Nov 24, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The guides may come apart during use, some guides may experience surface cracking, guide handles and modules may not target the femoral nail holes accurately and the instruments may not be able to be assemble together due to a change in the threading method used.
Source facts
- Action
- On July 1, 2009, U.S. distributors were notified by e-mail to return two instruments. The recall was expanded with international accounts notified by e-mail on August 5, 2009, and U.S. accounts being notified by e-mail on August 6, 2009, to return 7 instruments. The scope of the recall was then expanded again to include 14 instruments and another letter dated September 14, 2009 to distributors and user accounts. The notifications described the affected products, problem, health risks and actions for consignees. Direct questions about the recall to Zimmer, Inc. by calling 1-800-613-6131.
- Initiated
- 2009-07-01
- Root cause
- Other
- Terminated
- 2011-10-05
- Distribution
- Worldwide Distribution -- United States, France, Germany, Italy, Switzerland and United Kingdom.
- Recall number
- Z-0386-2010
- Recall status
- Terminated
- CDRH record ID
- 85323
- FDA classified
- 2009-11-24
- Recall event number
- 52806