Skip to content

Public FDA record

Z-0386-2022: activforce 2 package labeling: Model: Activ5-M R 201-200563

Recall record Z-0386-2022 for Activbody, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0386-2022
Company in FDA record
Activbody, Inc.
FDA record creation date
Dec 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to: 1) Manufacturing issue that result in inaccurate force measurement which may lead to may potential lead to incorrect treatment planning. 2) Firm failing to register and list its medical devices with U.S. Food and Drug Administration.

Source facts

Action
On 11/19/2021, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers and followed up with a hardcopy letter on 11/30/2021 via US Postal Service, to inform them that the firm failed to register with the U.S. Food and Drug Administration and therefore must be recalled from the market, and the risk of using the affected products can result in Inaccurate force measurements may result from using these devices. If inaccurate measurements are used to judge rehabilitation progress, the most immediate consequence is an increased risk of re-injury to the patient. Customers are instructed to: 1. Cease all use of these devices immediately. Use alternative methods and tools, such as manual muscle testing to measure force and/or a goniometer to measure range of motion. 2. Fill out the RECALL RETURN RESPONSE form. 3. Return the RECALL RETURN RESPONSE form and all devices in their possession to the Recalling Firm using the included pre-paid envelope. Customers only need to return the handheld dynamometer unit. Do not return the case, packaging or any accessories. For questions, contact support team via support@activbody.com
Initiated
2021-11-30
Root cause
Other
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, LA MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV. O.U.S.: Not provided
Recall number
Z-0386-2022
Recall status
Open, Classified
CDRH record ID
190596
FDA record created
2021-12-16
Recall event number
89070

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗