Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
Source facts
Action
Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
Initiated
2004-11-22
Root cause
Other
Terminated
2005-05-02
Distribution
Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
Recall number
Z-0387-05
Recall status
Terminated
CDRH record ID
36135
FDA classified
2005-01-08
Recall event number
30570