The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.
Source facts
Action
The firm sent letters with a return response form to customers. The firm instructed customers to immediately choose the "stop" command if presented with the 7-101 Liquid Sensor Error. Results from the assay should not be approved. A revision of the current software will have a mandatory installation soon.
Initiated
2017-07-05
Root cause
Device Design
Terminated
2018-10-04
Distribution
Nationwide
Recall number
Z-0387-2018
Recall status
Terminated
CDRH record ID
160281
FDA record created
2018-01-18
Recall event number
78752