Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.
Source facts
Action
The firm, GE Healthcare, sent an "Important Electronic Product Radiation Warning" letter to affected customers on September 10, 2018. The letter identified the affected product, problem and actions to be taken. GE Healthcare will, without charge, remedy the defect or bring the product into compliance. Your GE Service Representative will contact you to implement the correction.
For questions contact your local GEHC service representative or call 425-657-1512 or email: trisha.koenke@ge.com.
Initiated
2018-09-10
Root cause
Device Design
Terminated
2020-05-12
Distribution
Worldwide - US Nationwide Distribution
Recall number
Z-0388-2019
Recall status
Terminated
CDRH record ID
168762
FDA record created
2018-11-20
Recall event number
81454