Skip to content

Public FDA record

Z-0389-05: Zimmer™ Orthopaedic Surgical Products, Disposable Cuff, 12'' (30cm), Single Port, Single Bladder Sterile,

Recall record Z-0389-05 for Zimmer Orthopaedic Surgical Products. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0389-05
Company in FDA record
Zimmer Orthopaedic Surgical Products
FDA classification date
Jan 12, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The port/tube bond in the affected cuffs may separate during surgery causing the pressure in the cuff to release.

Source facts

Action
Consignees were first notified via certified mail on 10/27/2004. A follow up letter with 2 addtional lots were sent 11/11/2004.
Initiated
2004-10-27
Root cause
Other
Terminated
2005-12-20
Distribution
Nationwide, Australia, Demark, Germany
Recall number
Z-0389-05
Recall status
Terminated
CDRH record ID
35968
FDA classified
2005-01-12
Recall event number
30476

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗