Open original FDA record ↗Recall record
- FDA identifier
- Z-0389-2010
- Company in FDA record
- Smith & Nephew Inc
- FDA classification date
- Nov 20, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Radius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The deficiency may not be noticed by the user. Impingement and/or wear debris may lead to revision.
Source facts
- Action
- All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. Further questions are directed to the contacts at Smith and Nephew listed in the letter.
- Initiated
- 2009-01-23
- Root cause
- Other
- Terminated
- 2012-02-07
- Distribution
- Nationwide; 17 international: Switzerland, Dubai, Brazil, Canada, South Africa, Australia, Belgium, Spain, Thailand Turkey, Costa Rica, Denmark, India, Italy, and France.
- Recall number
- Z-0389-2010
- Recall status
- Terminated
- CDRH record ID
- 83472
- FDA classified
- 2009-11-20
- Recall event number
- 52615