Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue.
Source facts
Action
Mindray sent an Urgent Corrective Action letter dated October 7, 2013 to all of their customers. The letter identified the affected product, problem and actions to be taken. Customers should contact their Mindray Service Representative at 1-800-288-2121 Monday through Friday from 830 am to 530 pm to arrange for the replacement of the affected product.
Initiated
2013-10-07
Root cause
Environmental control
Terminated
2015-04-27
Distribution
Worldwide Distribution - USA Nationwide
Recall number
Z-0389-2014
Recall status
Terminated
CDRH record ID
123285
FDA classified
2013-11-25
Recall event number
66755