Z-0389-2022: Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system

Recall record Z-0389-2022 for Xstrahl Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0389-2022
Company in FDA record
Xstrahl Limited
FDA record creation date
Dec 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Base plate may detach from the main body of the treatment applicator.

Source facts

Action
The firm notified their consignees by email on 08/27/2021. The notice requested the following actions be taken: Xstrahl Limited are issuing the following inspection advice for all X150 specifical system users. 1. Prior to any clinical exposure the operator of the system should inspect the treatment applicator prior to inserting it into the treatment head of the X150. 2. As part of the daily assurance process the treatment applicators should be inspected. 3. If a treatment applicator is dropped it should be inspected by medical physics to ensure it is not damaged. 4. If the base plate is missing the applicators must be withdrawn from clinical use until it is repaired and accepted by medical physics. 5. In the case of an applicator where the base plate has become detached, review the prescription history for the patients for whom that applicator was prescribed. 6. Please advise Xstrahl Limited if you believe patients have been treated with a base plate missing, with detail of the applicator size and clinical filter, kV and HVL, and we will contact you directly.
Initiated
2021-08-27
Root cause
Device Design
Terminated
2026-03-04
Distribution
US and ALGERIA, AUSTRALIA, AUSTRIA, BULGARIA, CANADA, COSTA RICA, CROATIA, CYPRUS, DENMARK, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KAZAKHSTAN, LATVIA, MALTA, MEXICO, NEW ZEALAND, NORWAY, PANAMA, ROMANIA, RUSSIA, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UK, UKRAINE
Recall number
Z-0389-2022
Recall status
Terminated
CDRH record ID
188962
FDA record created
2021-12-16
Recall event number
88545

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗