Z-0392-05: Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System,

Recall record Z-0392-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0392-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Dec 30, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A software anomaly in the Logix CM software could result in a drug mixture being compounded without one of the ingredients if two drugs have the same indentifier in Logix and another order entry program being used with the Logix software.

Source facts

Action
Urgent Device Correction letters dated 12/8/04 were sent to the end users via overnight mail. The accounts were informed of the potential for omission of an ingredient if there are duplicate drug Id''s in the order, and were requested to complete and fax the customer reply form to Baxter at 847-270-5457, indicating if they are currently using Logix-CM software. Upon receipt of the reply form, Baxter will schedule a visit to the facility to examine the base and additive inventory information in their MOS database to confirm that there are no duplicate Drug ID numbers and that this issue has not occurred at the facility. They will also secure the MOS database to ensure that the situation will not be able to occur. Any questions were directed to Baxter Clintec Nutrition, option 1, at 1-800-422-2751.
Initiated
2004-12-08
Root cause
Other
Terminated
2005-11-01
Distribution
Maryland, Arizona, Indiana, Ohio, Minnesota, Pennsylvania and internationally to Canada
Recall number
Z-0392-05
Recall status
Terminated
CDRH record ID
36209
FDA classified
2004-12-30
Recall event number
30609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗