Open original FDA record ↗Recall record
- FDA identifier
- Z-0393-2022
- Company in FDA record
- Wright Medical Technology Inc
- FDA record creation date
- Dec 17, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The incorrect product is contained in the packaging.
Source facts
- Action
- The recalling firm issued a recall letter dated 03 Dec 2021 on 12/3/2021 via Fed Ex Priority Overnight. The letter explained the packages labeled as the CLAW II 4 Hole Plates and the packages labeled as the DARCO Locking Screws were switched, resulting in the packages containing the incorrect product. The letter also provided the potential risks, the actions needed to be taken, and provided example labels of affected products. The consignee is to immediately check their inventory to locate the packaged product and remove them from their point of use. The enclosed business reply form was to be returned via email to confirm receipt of the letter and document product segregation. After the firm receives the reply form, the consignee will be contacted to arrange for return and replacement. The consignee was given the choice of providing subdistribution contact details to the recalling firm so the recalling firm could inform them of the recall, or the consignee can notify their downstream customers of the recall.
- Initiated
- 2021-12-03
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution in the states of AZ, IA, ID, NC, NY, VA, and WV. The countries of Canada, France, Germany, Lebanon, The Netherlands, Saudi Arabia, and United Kingdom.
- Recall number
- Z-0393-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190462
- FDA record created
- 2021-12-17
- Recall event number
- 89089