Skip to content

Public FDA record

Z-0394-2022: WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Nu

Recall record Z-0394-2022 for Cardiovascular Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0394-2022
Company in FDA record
Cardiovascular Systems Inc
FDA classification date
Dec 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization.

Source facts

Action
Consignees were issued an URGENT: Medical Device Recall letter dated 11/22/2021. Consignees are asked to immediately remove from use and return affected devices to CSI. To do so: 1) Complete the attached Customer Acknowledgement Form. a) Per lot, document the quantity of all devices that are to be returned. b) Per lot, document the quantity of all devices that have been consumed. 2) Place the Acknowledgement Form with the device. 3) Return the device and Acknowledgement Form to CSI using the pre-paid label. A CSI Sales representative can return the device and form. Please contact CSI Recall Coordinator to organize pick up and return of the product. If all products specified in this recall notice have previously been consumed, used or are no longer available for return to CSI, please complete the Customer Acknowledgement Form per the above, and return the form to CSI in the provided envelope. For further info, please contact: Cardiovascular Systems, Inc. Field Action Hotline csifieldactionhotline@csi360.com 651-259-2800
Initiated
2021-11-22
Root cause
Device Design
Terminated
2023-08-31
Distribution
United States- AK, AL, AR, AZ, CA, DC, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV OUS - Netherlands
Recall number
Z-0394-2022
Recall status
Terminated
CDRH record ID
190574
FDA classified
2021-12-21
Recall event number
89120

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗