Skip to content

Public FDA record

Z-0394-2024: Intera 1.5T Power/Pulsar

Recall record Z-0394-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0394-2024
Company in FDA record
Philips North America Llc
FDA classification date
Nov 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential component failure in the Gradient Coil could product smoke and/or fire.

Source facts

Action
The firm mailed URGENT Medical Device Correction Notices dated 10/23/2023 to consignees to notify them of this recall event. Consignees are instructed to identify if they possess any affected devices in their inventory and to follow the provided instructions for when scan aborts occur to prevent the failure mode associated with this recall event. Consignees are to stop the scan immediately and contact Philips Service if any of the failure situations outlined in the notice occur during a scan. Consignees are asked to complete and return the provided response form to Philips no later than 30 days after receipt of the recall notice. Questions can be directed to Customer Care Solutions at 1-800-722-9377.
Initiated
2023-10-23
Root cause
Under Investigation by firm
Distribution
Domestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Luxembourg, Malaysia, Malta, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, United States, UAE, Venezuela, Vietnam, & Yemen.
Recall number
Z-0394-2024
Recall status
Open, Classified
CDRH record ID
204190
FDA classified
2023-11-21
Recall event number
93361

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗