Open original FDA record ↗Recall record
- FDA identifier
- Z-0395-05
- Company in FDA record
- Baxter Healthcare Renal Div
- FDA classification date
- Dec 2, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The air detector may not detect air bubbles consistently at the selected limit.
Source facts
- Action
- Important Product Information letters were sent to customers on 5/28/03, informing them of the potential for the air detector not detecting air bubbles consistently. The letters instructed the accounts to measure the air detector sensor voltage output on each of their instruments, using the enclosed instructions, to determine whether the installation of an air detector attenuator connector is needed. The instructions also included directions for installation of the attenuator connector. Since the instrument requires re-qualification after the testing and installation, the accounts were given the choice to have the testing and installation done during the next service call or as part of the instrument''s next normally scheduled maintenance.
- Initiated
- 2003-05-28
- Root cause
- Other
- Terminated
- 2008-02-26
- Distribution
- The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, the Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Korea, Lebanon, Malaysia, Morocco, the Netherlands, Oman, Palestine, Panama, Peru, the Philippines, Poland, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and the United Kingdom.
- Recall number
- Z-0395-05
- Recall status
- Terminated
- CDRH record ID
- 35989
- FDA classified
- 2004-12-02
- Recall event number
- 30486