Z-0395-2017: Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

Recall record Z-0395-2017 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0395-2017
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Nov 11, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded.

Source facts

Action
On 10/4/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Direct Consignees and Distributors via courier. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Direct Consignee Actions Required " Follow the instructions in the Customer Actions for Prevention/Detection of the Issue section of this Urgent Medical Device Correction (UMDC). " Complete the attached fax form (7184-00-1016) and fax it to 1-844-294-7181. " File this Urgent Medical Device Correction (UMDC) for future reference. Questions Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC. Distributor Actions Required " Ensure proper storage conditions (2-8C) are maintained for the reagents listed in the table above, per their respective method sheet. " If you have not directly shipped any of the products in the table above, you do not need to distribute the enclosed UMDC to your customers. " If you have directly shipped any of the products in the table above, please distribute the enclosed UMDC 16-157 to those customers. " Complete the attached fax form (7185-00-1016) and fax it to 1-844-294-7181. Questions Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this distributor letter or the enclosed UMDC 16-157.
Initiated
2016-10-04
Root cause
Storage
Terminated
2017-06-05
Distribution
Nationwide.
Recall number
Z-0395-2017
Recall status
Terminated
CDRH record ID
150149
FDA classified
2016-11-11
Recall event number
75366

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗