Open original FDA record ↗Recall record
- FDA identifier
- Z-0395-2018
- Company in FDA record
- Measurement Specialties Inc
- FDA record creation date
- Jan 19, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.
Source facts
- Action
- Customers were notified via mail or email on March 13, 2017. Additional customers were later identified and notified on June 14, 2017. Instructions included to disregard the sterilization instructions listed in the IFU (A08411F)originally provided with probes and instead utilize the updated IFU 08411G which contains validated sterilization instructions. They were also instructed to forward the recall notification to within their firm and to all organizations if further distributed. For further questions, please call (763) 316-4075.
- Initiated
- 2017-03-13
- Root cause
- Other
- Distribution
- USA (nationwide) distribution to the states of : AL, CA, KY, IL, KS, MN, NY, OH, TX, VA and WI. Distributed internationally to: Australia, Canada, China, England, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, Singapore, South Africa, South Korea, Taiwan, Vietnam
- Recall number
- Z-0395-2018
- Recall status
- Open, Classified
- CDRH record ID
- 158538
- FDA record created
- 2018-01-19
- Recall event number
- 78036