Skip to content

Public FDA record

Z-0396-2018: smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Platin

Recall record Z-0396-2018 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0396-2018
Company in FDA record
Smith & Nephew, Inc.
FDA record creation date
Jan 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa

Source facts

Action
The firm, Smith & Nephew, Inc., sent an "Urgent Medical Device Recall Notice" letter dated December 11, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect your inventory, locate unused devices and quarantine immediately; if you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out; return product, and complete and return the response form by email to FieldActions@smith-nephew.com or fax to 901-566-7975 even if you have not product to return. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com or call 901-566-7084.
Initiated
2017-12-11
Root cause
Mixed-up of materials/components
Terminated
2018-06-22
Distribution
US Distribution to states of: MI, GA, and OH.
Recall number
Z-0396-2018
Recall status
Terminated
CDRH record ID
160614
FDA record created
2018-01-19
Recall event number
78826

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗