Open original FDA record ↗Recall record
- FDA identifier
- Z-0396-2024
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Nov 21, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Poor image quality due to fluid ingress in the lens.
Source facts
- Action
- Notification of Urgent Medical Device Product Removal was issued via FedEx on Oct 3, 2023. The notice advises customers to immediately discontinue use and segregate affected product, and to complete and return the Reply Verification and Tracking Form (RVTF) even if you do not have any affected product. Customers are to post the notice near any affected product and notify end users. If further distributed, notify any customers to which product has been further distributed to. Reach out to your local Boston Scientific representative with any questions. The notice is batch / lot specific. Boston Scientific has implemented corrective changes and will make replacement EXALT Model D scopes available as soon as possible.
- Initiated
- 2023-09-27
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA VT, WA, WI & WV. The countries of Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Great Britain, Italy, New Zealand, Spain, Sweden, Switzerland.
- Recall number
- Z-0396-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203827
- FDA classified
- 2023-11-21
- Recall event number
- 93281