Open original FDA record ↗Recall record
- FDA identifier
- Z-0397-06
- Company in FDA record
- Plus Orthopedics USA
- FDA classification date
- Jan 18, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
Source facts
- Action
- The adjustment know tower on one device was found t have broken off from the body of the instrument. This did not happend during a surgery, and is belived tohave been caused by the instrument being dropped or otherwise mishandled during hospital processing. A manufacturer evaluation of the device determined that the laser weld attaching the knob to the body was not to specification.
- Initiated
- 2005-12-02
- Root cause
- Other
- Terminated
- 2007-01-29
- Distribution
- Arizona, California, Colorado, Idaho, New York, Oklahoma, Texas, Utah, Wisconsin
- Recall number
- Z-0397-06
- Recall status
- Terminated
- CDRH record ID
- 43345
- FDA classified
- 2006-01-18
- Recall event number
- 34255