Skip to content

Public FDA record

Z-0397-2016: IBS - Osteosynthesis screws for bone fixation Model S25 ST014.

Recall record Z-0397-2016 for In2Bones, SAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0397-2016
Company in FDA record
In2Bones, SAS
FDA classification date
Dec 9, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One batch of osteosynthesis screws were manufactured with the incorrect Torx imprint on the screw head.

Source facts

Action
IN2Bones - USA sent an Field Safety Corrective Action Medical Device Batch Recall letter dated November 17, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Recommended actions : We therefore you to follow the instructions here below: 1. Identify all I.B.screws with the batches subject of this recall that might still be in your inventory and quarantine them 2. Inform and distribute this Recall Notification to all relevant persons within your organization 3. To distributors only : Identify all I.B.screws with the batches subject of this recall that were delivered to your customers, and if relevant instruct them to also follow these instructions (identification and quarantine) 4. Fill in and return the fax back form enclosed. With this form, you will certify that you have received this Recall Notification and intend to, comply with the recommendations listed.This fax back form will enable ln2Bones to conduct effectiveness checks. For any question,plea1se contact our Quality and Regulatory Affairs team at: +33 4 72 29 26 26/ +33 7 6188 44 30 or by email:qualite@in2bones.com.
Initiated
2015-11-18
Root cause
Nonconforming Material/Component
Terminated
2016-08-19
Distribution
US Distribution to the state of :TN
Recall number
Z-0397-2016
Recall status
Terminated
CDRH record ID
142015
FDA classified
2015-12-09
Recall event number
72711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗