Skip to content

Public FDA record

Z-0397-2024: EXALT Model D Single-Use Duodenoscope, 2 pack: Outer box UPN# M00542421, Inner box UPN # M00542420 & Oute

Recall record Z-0397-2024 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0397-2024
Company in FDA record
Boston Scientific Corporation
FDA classification date
Nov 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Poor image quality due to fluid ingress in the lens.

Source facts

Action
Notification of Urgent Medical Device Product Removal was issued via FedEx on Oct 3, 2023. The notice advises customers to immediately discontinue use and segregate affected product, and to complete and return the Reply Verification and Tracking Form (RVTF) even if you do not have any affected product. Customers are to post the notice near any affected product and notify end users. If further distributed, notify any customers to which product has been further distributed to. Reach out to your local Boston Scientific representative with any questions. The notice is batch / lot specific. Boston Scientific has implemented corrective changes and will make replacement EXALT Model D scopes available as soon as possible.
Initiated
2023-09-27
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA VT, WA, WI & WV. The countries of Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Great Britain, Italy, New Zealand, Spain, Sweden, Switzerland.
Recall number
Z-0397-2024
Recall status
Open, Classified
CDRH record ID
203828
FDA classified
2023-11-21
Recall event number
93281

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗