Skip to content

Public FDA record

Z-0398-06: ARCHITECT c8000 System Software, Version 2.11, Product List/Model #5F48-12, distributed by Abbot Laborato

Recall record Z-0398-06 for Abbott Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0398-06
Company in FDA record
Abbott Laboratories, Inc
FDA classification date
Jan 18, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL.

Source facts

Action
Firm initiated recall to all consignees via letter dated 12/08/05.
Initiated
2005-12-08
Root cause
Other
Terminated
2007-09-25
Distribution
Worldwide. Product distributed nationwide to direct accounts (including gov''t accounts) in: AL, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MN, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WY. Foreign distribution to the following countries: Argentina, Australia, Bermuda, Brazil, Canada, Chile, Columbia, Ecuador, El Salvador, Germany, Grand Cayman, Guatemala, Hong Kong, Jamaica, Mexico, New Zealand, Panama, Puerto Rico, Singapore, Thailand, Trinidad, Uruquay and Venezuela.
Recall number
Z-0398-06
Recall status
Terminated
CDRH record ID
43391
FDA classified
2006-01-18
Recall event number
34266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗