Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device).
Source facts
Action
All associated LOTs shall be accounted for and all devices in those lots. Since only two LOTs are affected and only two costumers/distributors are in possession of said LOTs no failed attempt to correspond with them is expected. If any of the distributors fails to respond, a Gordian costumer representative will personally go to location to follow through on the process indicated.
Initiated
2020-09-22
Root cause
Process control
Terminated
2021-02-11
Distribution
US Nationwide distribution including in the states of Texas and Michigan.
Recall number
Z-0398-2021
Recall status
Terminated
CDRH record ID
183815
FDA record created
2020-11-03
Recall event number
86483