- FDA identifier
- Z-0399-05
- Company in FDA record
- Cardiac Science, Inc.
- FDA classification date
- Jan 15, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the event that charge noise is presrent it will delay the first defibrillation shock in the 3 shock sequence.
Source facts
- Action
- Cardiac Science will contact customers via telephone to set up a time for Clinical Educator to visit the customer and conduct a field test on each AED. AEDs that fail the field test will be replaced either on the spot or within 24 hours. AEDs failing the test are either removed by the Clinical Educator or are returned to Cardiac Science upon receiving the new AED. Cardiac Science anticipates testing to be done by the end of December 2004.
- Initiated
- 2004-11-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2006-06-25
- Distribution
- Throughout the United States, Canada and Europe.
- Recall number
- Z-0399-05
- Recall status
- Terminated
- CDRH record ID
- 36660
- FDA classified
- 2005-01-15
- Recall event number
- 30704