Skip to content

Public FDA record

Z-0399-05: Powerheart AED G3 Automated External Defibrillator Model 9300A (Automatic). Cardiac Science AEDs are man

Recall record Z-0399-05 for Cardiac Science, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0399-05
Company in FDA record
Cardiac Science, Inc.
FDA classification date
Jan 15, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the event that charge noise is presrent it will delay the first defibrillation shock in the 3 shock sequence.

Source facts

Action
Cardiac Science will contact customers via telephone to set up a time for Clinical Educator to visit the customer and conduct a field test on each AED. AEDs that fail the field test will be replaced either on the spot or within 24 hours. AEDs failing the test are either removed by the Clinical Educator or are returned to Cardiac Science upon receiving the new AED. Cardiac Science anticipates testing to be done by the end of December 2004.
Initiated
2004-11-12
Root cause
Nonconforming Material/Component
Terminated
2006-06-25
Distribution
Throughout the United States, Canada and Europe.
Recall number
Z-0399-05
Recall status
Terminated
CDRH record ID
36660
FDA classified
2005-01-15
Recall event number
30704

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗