Skip to content

Public FDA record

Z-0400-2014: LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). M

Recall record Z-0400-2014 for Lanx, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0400-2014
Company in FDA record
Lanx, Inc.
FDA classification date
Nov 26, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.

Source facts

Action
LANX sent an URGENT: Recall letter dated November 5, 2013, to all affected customers via Federal Express, Overnight delivery. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return all affected product to the firm. Customers with questions were instructed to contact Field Inventory Management. For questions regarding this recall call 303-501-8402.
Initiated
2013-11-04
Root cause
Process control
Terminated
2015-12-29
Distribution
Nationwide Distribution including AZ, PA, WI, NY, TX, PR, and TN.
Recall number
Z-0400-2014
Recall status
Terminated
CDRH record ID
123690
FDA classified
2013-11-26
Recall event number
66836

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗